multi N/C 2100S Fault removal
95
80 no connection to the temperature controller
Cause Remedy
no connection to the HT furnace solids
module
not switched on
incorrect connection
switch on solids module
check connection cable
check connection
81 Thermocouple HT furnace interruption
Cause Remedy
faulty thermocouple
furnace not connected
furnace temperature too high
inform Service
Connect furnace
inform Service
84 Communication error HT furnace temperature controller
Cause Remedy
Communication errors inform Service
86 No external furnace available
Cause Remedy
no connection to the solids module check connection cable
113 AS 60: z drive error (steps lost)
Cause Remedy
sampler AS 60 misaligned in z direction
(immersion depth)
check the sampler adjustment
200 Restart computer in the analysis device
Cause Remedy
internal computer reset
overvoltage
short-term power failure
if the front LED illuminates, initialize the analyzer
for repeated occurrence monitor precisely at which
time the error occurs (note status line)
201 Restart the internal program
Cause Remedy
internal program error initialize analyzer
for repeated occurrence monitor precisely at which
time the error occurs (note status line)
MESSx Analyzer error: MESSx measurement is cancelled
Cause Remedy
Equipment fault initialize analyzer
check window SYSTEM STATE after error removal
search for equipment fault and remove error
Peltier temperature outside range
Cause Remedy
Peltier cooling insufficient
Inform Service. After repair, replace water traps.