EasyManua.ls Logo

Endress+Hauser MODBUS RS485 - Page 114

Endress+Hauser MODBUS RS485
140 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Troubleshooting Proline Promag 53 MODBUS RS485
114 Endress+Hauser
Removing the electrode Fitting the electrode
1 Release the socket head cap screw (1) and remove
the cap.
1 Insert the new electrode (5) from underneath into
the retaining cylinder (7). Ensure that the seals at
the tip of the electrode are clean.
2 Unscrew the electrode cable (3) secured on the
rotary handle (2).
2 Attach the rotary handle (2) to the electrode and
secure with the locking bolt (8).
"
Caution!
Ensure that the coil spring (11) is inserted to
guarantee perfect electrical contact and thus correct
measuring signals.
3 Release the knurled nut (4) by hand.
This knurled nut serves as a counter nut.
3 Pull the electrode back until the tip is no longer
protruding from the retaining cylinder (7).
4 Unscrew the electrode (5) using the rotary handle
(2). It can now be pulled out of the retaining
cylinder (7) up to a defined point.
#
Warning!
Risk of injury!
The electrode can bounce back to its stop under
process conditions (pressure in the pipe). Exert
counter pressure against it when releasing.
4 Screw the retaining cylinder (7) onto the ball valve
housing (9) and secure tightly by hand.
The seal (10) at the retaining cylinder must be fitted
and clean.
!
Note!
Ensure that the rubber hoses attached to the
retaining cylinder (7) and stop cock (6) are the same
color (red or blue).
5 Close the stop cock (6) after you have pulled out the
electrode as far as it will go.
#
Warning!
Do not open the stop cock after this to ensure that
no fluid comes out.
5 Open the stop cock (6) and screw the electrode as
far as it will go into the retaining cylinder using the
rotary handle (2).
6 You can now unscrew the entire electrode with the
retaining cylinder (7).
6 Now, screw the knurled nut (4) on the retaining
cylinder. This secures the electrode in position.
7 Remove the rotary handle (2) from the electrode (5)
by pushing out the locking bolt (8). Ensure that you
do not lose the coil spring (11).
7 Secure the electrode cable (3) on the rotary handle
(2) once more using the socket head cap screw.
"
Caution!
Ensure that the socket head cap screw of the
electrode cable is securely tightened. This
guarantees perfect electrical contact and thus correct
measuring signals.
8 Now replace the old electrode with the new one.
You can order replacement electrodes separately
from Endress+Hauser
8 Refit the cover and tighten the socket head cap
screw (a).

Table of Contents

Related product manuals