OXY5500 Optical Oxygen Analyzer BA02195C
5
–10 Endress+Hauser
Table 5-1 Holding registers (Continued)
Register
Name
Address Size
Variable
Type
Description
Write
Access
Cal0 5521 2 Float Calibration value: Phase
shift of low oxygen calibra-
tion point (Default: 59.9).
Yes
T0 5523 2 Float Calibration value: Temper-
ate at the low oxygen cali-
bration point in
° C
(Default: 20.0).
Yes
O2-2nd 5527 2 Float Calibration value: Oxygen
concentration of the high
oxygen calibration point in
the unit defined in register
5535 (O2-2nd
Unit).
Yes
Cal-2nd 5529 2 Float Calibration value: Phase
shift of high oxygen cali-
bration point (Default:
26.3).
Yes
T2nd 5531 2 Float Calibration value: Tem-
perature at the high cali-
bration point in
° C.
Yes
pATM 5533 2 Float Calibration value: Pressure
at the high oxygen calibra-
tion in mbar.
Yes
O2-2nd Unit 5535 2 Int32 The unit for the O2-2nd
value, where:
0x4000.0000 = ppmv
0x0000.0010 = % O
2
Yes
Ethernet
Obtain IP
Mode
5675 2 Int32 Activates or deactivates
DHCP. Entering “1” will
obtain IP automatically.
Yes
Ethernet IP 5677 8 Int32 The Ethernet IP. Each pair
of registers holds one octet
of the address. Refer to
“Ethernet IP” on page
5-17 for details. This regis-
ter will only be used if reg-
ister 5675 is set to “0”
(DHCP off).
Yes