BA02195C OXY5500 Optical Oxygen Analyzer
Endress+Hauser 5
9
4. Changing the sensor type will not change any other sensor constraint.
Table 5-1 Holding registers (Continued)
Register
Name
Address Size
Variable
Type
Description
Write
Access
4-20mA
Input
Calibration
Values
5667 8 Float Two calibration values for a
low point and a high point,
each with reference value
and device output. Refer to
“4-20mA input calibra-
tion values” on page
5-16.
Yes
Measure-
ment Values
4895 14 Mixed Refer to “Measurement
values” on page 5-14 for
details.
No
Sensor
Constant f1
4911 2 Float Sensor constant f1.
Allowed range: 0.000 to
9.999
Yes
Sensor
Constant
dPhi1
4913 2 Float Sensor constant dPhi1.
Allowed range: -9.99999 to
+9.99999
Yes
Sensor
Constant
dPhi2
4917 2 Float Sensor constant dPhi2.
Allowed range: -9.99999 to
+9.99999
Yes
Sensor
Constant
dKSV1
4919 2 Float Sensor constant dKSV1.
Allowed range: -9.99999 to
+9.99999
Yes
Sensor
constant
DKSV2
4921 2 Float Sensor constant dKSV2.
Allowed range: -9.99999 to
+9.99999
Yes
Sensor
Constant m
4923 2 Float Sensor constant m.
Allowed range: 0.00 to
+999.99
Yes
Sensor Type 4925 2 Int32 Sensor Type where:
0x00 = OP-3
0x01 = OP-6
0x02 = OP-9
4
Yes
Manual
Temperature
Compensa-
tion
5611 1 Boolean Activates the temperature
measurement of the Pt100
sensor by setting and uses
the manual temperature
value by deleting this bool-
ean register. After writing
this register, the manual
temperature value has to
be set (register 2411).
Yes