OXY5500 Optical Oxygen Analyzer BA02195C
5
–8 Endress+Hauser
3. This value cannot be changed.
Table 5-1 Holding registers (Continued)
Register
Name
Address Size
Variable
Type
Description
Write
Access
4-20mA
Port2 Error
Trigger Level
Value
4975 2 Int32 Output current in the event
of an error, where:
0x00 = 22mA
0x01 = 2mA
Yes
4-20mA
Port2
Calibration
Values
4979 8 Float Two calibration values for a
low point and a high point,
each with a reference value
and device output. Refer to
“4-20mA port2 calibra-
tion values” on page
5-16.
Yes
4-20mA
Input
Interface
5633 2 Int32 This register is reserved for
future use.
Yes
4-20mA
Input
Channel
5637 2 Int32 Code for 4-20mA Port1
Output Interface where:
0x02 = Pressure.
3
No
4-20mA
Input Low
Value
5651 2 Float The 4mA corresponding
input value. Refer to “Ana-
log input and output
value ranges” on page
5-16 for details.
Yes
4-20mA
Input Mid
Value
5653 2 Float The 12mA input value is
only used in bilinear mode.
Refer to “Analog input
and output value
ranges” on page 5-16 for
details.
Yes
4-20mA
Input High
Value
5655 2 Float The 20mA input value.
Refer to “Analog input
and output value
ranges” on page 5-16 for
details.
Yes
4-20mA
Input Fixed
Value
5657 2 Float This register is reserved for
future use.
Yes
4-20mA
Input Error
Trigger Level
Value
5663 2 Float This register is reserved for
future use.
Yes