Proline Prowirl F 200 PROFIBUS PA Commissioning
Endress+Hauser 115
Parameter Prerequisite Description Selection / User
entry
Factory setting
Pressure unit For the following order code:
"Sensor version", option "Mass
flow"
Select process pressure unit.
Effect
The unit is taken from the:
• Calculated saturated steam
pressure
• Atmospheric pressure
• Maximum value
• Fixed process pressure
• Pressure
• Reference pressure
Unit choose list Country-specific:
• bar
• psi
Velocity unit – Select velocity unit.
Result
The selected unit applies for:
• Flow velocity
• Maximum value
Unit choose list Country-specific:
• m/s
• ft/s
Unit totalizer One of the following options is
selected in the Assign process
variable parameter of the
Totalizer 1 to 3 submenu:
• Volume flow
• Corrected volume flow
• Mass flow
• Total mass flow
*
• Condensate mass flow
*
• Energy flow
*
• Heat flow difference
*
Select process variable totalizer
unit.
Unit choose list Country-specific:
• m³
• ft³
Temperature unit – Select temperature unit.
Effect
The selected unit applies for:
• Temperature
• Maximum value
• Minimum value
• Average value
• Maximum value
• Minimum value
• Maximum value
• Minimum value
• 2nd temperature delta heat
• Fixed temperature
• Reference combustion
temperature
• Reference temperature
• Saturation temperature
Unit choose list Country-specific:
• °C
• °F
Switch-on value • The Switch option is
selected in the Operating
mode parameter.
• The Limit option is selected
in the Switch output
function parameter.
Enter measured value for the
switch-on point.
Signed floating-point
number
Country-specific:
• 0 m³/h
• 0 ft³/h
Switch-off value • The Switch option is
selected in the Operating
mode parameter.
• The Limit option is selected
in the Switch output
function parameter.
Enter measured value for the
switch-off point.
Signed floating-point
number
Country-specific:
• 0 m³/h
• 0 ft³/h
Switch-on delay • The Switch option is
selected in the Operating
mode parameter.
• The Limit option is selected
in the Switch output
function parameter.
Define delay for the switch-on
of status output.
0.0 to 100.0 s 0.0 s