Proline Prowirl F 200 PROFIBUS PA Diagnostics and troubleshooting
Endress+Hauser 167
Diagnostic information Remedy instructions Influenced measured
variables
No. Short text
870 Measuring inaccuracy increased 1. Check process
2. Increase flow volume
Calculated saturated
steam pressure
Density
Energy flow
Flow velocity
Heat flow difference
Low flow cut off
Mass flow
Total mass flow
Switch output status
Pressure
Reynolds number
Specific volume
Corrected volume flow
Steam quality
Degrees of superheat
Volume flow
Measured variable status [from the factory]
1)
Quality Uncertain
Quality substatus Process related
Coding (hex) 0x78 to 0x7B
Status signal S
Diagnostic behavior Warning
1) Diagnostic behavior can be changed. This causes the overall status of the measured variable to change.
Diagnostic information Remedy instructions Influenced measured
variables
No. Short text
871 Near steam saturation limit Check process conditions Calculated saturated
steam pressure
Density
Energy flow
Heat flow difference
Low flow cut off
Mass flow
Total mass flow
Switch output status
Pressure
Reynolds number
Specific volume
Corrected volume flow
Steam quality
Degrees of superheat
Measured variable status [from the factory]
1)
Quality Uncertain
Quality substatus Process related
Coding (hex) 0x78 to 0x7B
Status signal S
Diagnostic behavior Warning
1) Diagnostic behavior can be changed. This causes the overall status of the measured variable to change.
Diagnostic information Remedy instructions Influenced measured
variables
No. Short text
872 Wet steam detected 1. Check process
2. Check plant
Energy flow
Heat flow difference
Low flow cut off
Total mass flow
Switch output status
Corrected volume flow
Steam quality
Measured variable status [from the factory]
1)
Quality Uncertain
Quality substatus Process related
Coding (hex) 0x78 to 0x7B
Status signal S
Diagnostic behavior Warning
1) Diagnostic behavior can be changed. This causes the overall status of the measured variable to change.