Commissioning Proline Promag H 300 PROFINET
102 Endress+Hauser
Parameter overview with brief description
Parameter Prerequisite Description Selection / User
interface / User
entry
Factory setting
Relay output function Select the function for the
relay output.
Closed
Open
Diagnostic
behavior
Limit
Flow direction
check
Digital Output
Closed
Terminal number Shows the terminal numbers
used by the relay output
module.
Not used
24-25 (I/O 2)
22-23 (I/O 3)
Assign flow direction check In the Relay output function
parameter, the Flow direction
check option is selected.
Select process variable for flow
direction monitoring.
Off
Volume flow
Mass flow
Corrected volume
flow
Volume flow
Assign limit The Limit option is selected in
the Relay output function
parameter.
Select process variable for limit
function.
Off
Volume flow
Mass flow
Corrected volume
flow
Flow velocity
Conductivity
*
Corrected
conductivity
*
Totalizer 1
Totalizer 2
Totalizer 3
Temperature
*
Electronic
temperature
Volume flow
Assign diagnostic behavior In the Relay output function
parameter, the Diagnostic
behavior option is selected.
Select diagnostic behavior for
switch output.
Alarm
Alarm or warning
Warning
Alarm
Assign status In the Relay output function
parameter, the Digital Output
option is selected.
Select device status for switch
output.
Partially filled pipe
detection
Low flow cut off
Profinet Slot 18
*
Profinet Slot 19
*
Profinet Slot 20
*
Partially filled pipe
detection
Switch-off value In the Relay output function
parameter, the Limit option is
selected.
Enter measured value for the
switch-off point.
Signed floating-point
number
Country-specific:
0 l/h
0 gal(us)/min
Switch-off delay In the Relay output function
parameter, the Limit option is
selected.
Define delay for the switch-off
of status output.
0.0 to 100.0 s 0.0 s
Switch-on value The Limit option is selected in
the Relay output function
parameter.
Enter measured value for the
switch-on point.
Signed floating-point
number
Country-specific:
0 l/h
0 gal(us)/min
Switch-on delay In the Relay output function
parameter, the Limit option is
selected.
Define delay for the switch-on
of status output.
0.0 to 100.0 s 0.0 s
Failure mode Define output behavior in
alarm condition.
Actual status
Open
Closed
Open
* Visibility depends on order options or device settings