Proline Promag H 300 PROFINET Commissioning
Endress+Hauser 107
Parameter overview with brief description
Parameter Prerequisite Description Selection / User
entry
Factory setting
Assign process variable Select process variable for low
flow cut off.
Off
Volume flow
Mass flow
Corrected volume
flow
Volume flow
On value low flow cutoff A process variable is selected
in the Assign process variable
parameter (→  107).
Enter on value for low flow cut
off.
Positive floating-
point number
Depends on country
and nominal
diameter
Off value low flow cutoff A process variable is selected
in the Assign process variable
parameter (→  107).
Enter off value for low flow cut
off.
0 to 100.0 % 50 %
Pressure shock suppression A process variable is selected
in the Assign process variable
parameter (→  107).
Enter time frame for signal
suppression (= active pressure
shock suppression).
0 to 100 s 0 s