Proline Promag H 300 PROFINET Diagnostics and troubleshooting
Endress+Hauser 165
Diagnostic information Remedy instructions Influenced measured
variables
No. Short text
437 Configuration incompatible 1. Restart device
2. Contact service
• Conductivity
• Corrected conductivity
• Measured values 1
• Measured values 2
• Measured values 3
• Density
• Electronic temperature
• Flow velocity
• Mass flow
• Coil current shot time
• Reference electrode
potential against PE
• Noise
• Reference density
• Corrected volume flow
• Temperature
• Status
• Volume flow
Measured variable status
Quality Bad
Quality substatus Maintenance alarm
Coding (hex) 0x24 to 0x27
Status signal F
Diagnostic behavior Alarm
Diagnostic information Remedy instructions Influenced measured
variables
No. Short text
438 Dataset 1. Check data set file
2. Check device configuration
3. Up- and download new configuration
• Conductivity
• Corrected conductivity
• Measured values 1
• Measured values 2
• Measured values 3
• Density
• Electronic temperature
• Flow velocity
• Mass flow
• Coil current shot time
• Reference electrode
potential against PE
• Noise
• Reference density
• Corrected volume flow
• Temperature
• Status
• Volume flow
Measured variable status
Quality Uncertain
Quality substatus Maintenance demanded
Coding (hex) 0x68 to 0x6B
Status signal M
Diagnostic behavior Warning
Diagnostic information Remedy instructions Influenced measured
variables
No. Short text
441 Current output 1 to n 1. Check process
2. Check current output settings
–
Measured variable status [from the factory]
1)
Quality Good
Quality substatus Ok
Coding (hex) 0x80 to 0x83
Status signal S
Diagnostic behavior Warning
1) Diagnostic behavior can be changed. This causes the overall status of the measured variable to change.