Proline Promag H 300 PROFINET System integration
Endress+Hauser 79
Available units
Density Temperature
Unit code Unit Unit code Unit
1100 g/cm
3
1001 °C
1101 g/m
3
1002 °F
1099 kg/dm
3
1000 K
1103 kg/l 1003 °R
1097 kg/m
3
1628 SD4°C
1629 SD15°C
1630 SD20°C
32833 SG4°C
32832 SG15°C
32831 SG20°C
1107 lb/ft
3
1108 lb/gal (us)
32836 lb/bbl (us;liq.)
32835 lb/bbl (us;beer)
32837 lb/bbl (us;oil)
32834 lb/bbl (us;tank)
1403 lb/gal (imp)
32838 lb/bbl (imp;beer)
32839 lb/bbl (imp;oil)
Data structure
Output data of Analog Output
Byte 1 Byte 2 Byte 3 Byte 4 Byte 5 Byte 6 Byte 7
Measured value: floating point number (IEEE 754) Status
1)
Unit code
1) Status coding → 81
Failsafe mode
A failsafe mode can be defined for using the compensation values.
If the status is GOOD or UNCERTAIN, the compensation values transmitted by the
automation system are used. If the status is BAD, the failsafe mode is activated for the use
of the compensation values.
Parameters are available per compensation value to define the failsafe mode: Expert
→ Sensor → External compensation
Fail safe type parameter
• Fail safe value option: The value defined in the Fail safe value parameter is used.
• Fallback value option: The last valid value is used.
• Off option: The failsafe mode is disabled.
Fail safe value parameter
Use this parameter to enter the compensation value which is used if the Fail safe value
option is selected in the Fail safe type parameter.