Operation
Eppendorf Xplorer
®
/Eppendorf Xplorer
®
plus
English (EN)
37
5.11 Dispensing of liquid – Dis selection dial setting
In the Dispensing (Dis) mode, liquid is aspirated and individually dispensed in uniform
partial volumes. The reverse stroke and remaining stroke are triggered before or after the
dispensing steps.
Prerequisites
The pipette is switched on.
The Dispensing (Dis) operating mode has been selected.
The parameters have been set.
The pipette tip has been attached.
1. Press the rocker up to aspirate liquid.
2. Press the rocker down to execute the reverse stroke.
3. Press the rocker down again for each dispensing operation. The display indicates the
remaining dispensing steps under Steps.
After dispensing, decide whether the residual liquid should be dispensed or liquid should
be aspirated again.
Liquid aspiration: press the rocker up once.
Liquid dispensing: press the rocker down twice.
1 Aspiration
2 Reverse stroke
3 Dispensing
4 Aspirate or empty
Press the Empty softkey to cancel dispensing. The tip is completely emptied.
Press the rocker up to stop a current dispensing of a large dispensing volume.
1234 6
5
1 2 3 4