Operation
Eppendorf Xplorer
®
/Eppendorf Xplorer
®
plus
English (EN)
For the Set interval parameter, set the required interval. If you do not want to use the
Service interval option, set the value of the parameter to 0. Triggering the reset function
resets the interval. If the pipette should start counting at a specific initial value, enter the
value under Set interval and then select Reset. You can view the last time a reset was
triggered for the executed maintenance under Last service. You can view when the next
maintenance needs to be executed under Next service. If the interval has elapsed up to
90%, the pipette shows the next required maintenance in the display. Monitoring of the
interval time and the interval cycles function independently of each other. If you activate
the interval time and interval cycles, you will receive a maintenance reminder for both
settings. A reset may be required after the maintenance and the gravimetric test.
5.19.14 Setting the Date and Time (Xplorer plus)
Prerequisites
The pipette is switched on.
The Opt operating mode has been selected.
The pipette is switched on.
Select Date or Time using the rocker.
Press the Select softkey.
Change the value using the rocker.
Press the Next softkey to change the field.
Press the End softkey to end the entry.
Change the second parameter in the same way.
Press the End softkey to end the entry.
5.20 Resetting the pipette to the factory setting
All settings (language, selected parameters, options, etc.) of the pipette are reset to their
initial state. All the saved program and adjustment settings are deleted.
The Initial reset function also overwrites existing password protection.
5.20.1 Performing Initial reset – Opt selection dial setting
Prerequisites
The pipette is switched on.
The Option operating mode has been selected.
1. Select Service from the list.
2. Select Initial reset from the list.
3. Confirm the confirmation prompt.
The pipette is reset to its initial state.