5.15.3 Sequential dispensing of liquid – Spc selection dial setting
Up to 10 different volumes can be dispensed in the Seq. Dispensing mode. The sum of the
dispensing volumes may not exceed the pipette tip filling volume. This mode is well
suited for dilution series. The Seq. Pipetting mode is also well suited for dilution series. To
conduct a dilution series, dispense two different liquids in the opposite direction. This
allows you to achieve different dilutions with the same volume per tube.
Prerequisites
The pipette is switched on.
The Seq. Dispensing (Spc) operating mode has been selected.
The parameters have been set
The pipette tip has been attached.
1. Press the rocker up to aspirate liquid.
2. Press the rocker down to execute the reverse stroke.
3. Press the rocker down for each dispensing step.
4. If the piston is in the basic position: press the Back softkey to exit the mode.
If the sum of the dispensing volumes exceeds the volume of the pipette tip:
create a program and execute the mode several times in a row.
1 Sequential dispensing of solution 1 2 Sequential dispensing of solution 2 in
the opposite direction
1234 6
5
1234 6
5
1 2