5.15.2.1 Examples:
Required auxiliary aids:
• Xplorer plus with a nominal volume of 1 000 µL
• Dispensing tube (1.5 mL)
Example 1 - total volume 500 μL, dilution 1:10
• Diluent: 450 µL
• Air bubble: 300 µL - 500 µL
• Sample or reagent: 50 µL
Example 2 - total volume 900 μL, dilution 1:10
• Diluent: 810 µL
• Air bubble: 100 µL
• Sample or reagent: 90 µL
Example 3 – total volume 1 000 μL, dilution 1:10
• Diluent: 900 µL
• Air bubble: 0 µL
• Sample or reagent: 100 µL