1. Press the rocker up to aspirate the blow-out and the liquid to be dispensed.
The aspiration of the dispensing volume and blow-out will be displayed under the
dispensing volume with the label + Blow .
2. Press the rocker down to dispense the liquid.
The piston returns to the basic position. Liquid remains in the tip after the dispensing
volume has been dispensed.
• Press the rocker up to aspirate the same liquid.
• Press the rocker down to end the pipetting and discard the blow-out (Blow).
5.15.5 Sequential pipetting of liquid – Spc selection dial setting
In the Seq. Pipetting mode, a maximum of 10 different pipetting volumes are dispensed in
a fixed sequence. This mode is well suited for dilution series. The Seq. Dispensing mode is
also well suited for dilution series. The random error (precision) and systematic error
(accuracy) are better for pipetting than with dispensing. The Seq. Pipetting mode offers
more flexibility with the volume selection than the Seq. Dispensing mode.
Prerequisites
• The pipette is switched on.
• The Seq. Pipetting (Spc) operating mode has been selected.
• The parameters have been set.
• The pipette tip has been attached.
1. Press the rocker up to aspirate liquid.
2. Press the rocker down to dispense the liquid.
3. Press the rocker downward after the dispensing to execute the blow-out (Blow).
4. If the piston is in the basic position: press the Back softkey to exit the mode.
If necessary, activate the Counter setting in the options to count the executed
dispensings using the pedometer.
1 Aspiration
2 Dispensing
3 Aspiration
4 Dispensing
1 3 42