8.2 Measurement errors of the multi-channel pipette
The measurement errors apply to 8-channel pipettes and 12-channel pipettes.
8.3 Test conditions
Test conditions and test analysis in accordance with the standard ISO 8655, Part 6. The
inspections are performed with a calibrated analytical balance with evaporation
protection.
Number of determinations per volume: 10
Water according to ISO 3696
Inspection at 20 °C – 25 °C ±0.5 °C
Dispensing onto the tube inner wall
Operating mode: PIP
Speed level: 5
Model Test tip
epT.I.P.S.
Testing
volume
Error of measurement
Systematic error Random error
± μL ± μL
0.5 µL10 µL 0.1 µL 20 µL
medium gray
40 mm
1 µL 5.0 0.05 3.0 0.03
5 µL 3.0 0.15 1.5 0.075
10 µL 2.0 0.2 0.8 0.08
5 µL100 µL 2 µL 200 µL
yellow
53 mm
10 µL 2.0 0.2 2.0 0.2
50 µL 1.0 0.5 0.8 0.4
100 µL 0.8 0.8 0.25 0.25
15 µL300 µL 15 µL 300 µL
orange
55 mm
30 µL 2.5 0.75 1.0 0.3
150 µL 1.0 1.5 0.5 0.75
300 µL 0.6 1.8 0.25 0.75
50 µL1200 µL 50 µL 1250 µL
green
76 mm
120 µL 6.0 7.2 0.9 1.08
600 µL 2.7 16.2 0.4 2.4
1200 µL 1.2 14.4 0.3 3.6