8.2 Measurement errors of the multi-channel pipette
The measurement errors apply to 8-channel pipettes and 12-channel pipettes.
8.3 Test conditions
Test conditions and test analysis in accordance with the standard ISO 8655, Part 6. The
inspections are performed with a calibrated analytical balance with evaporation
protection.
• Number of determinations per volume: 10
• Water according to ISO 3696
• Inspection at 20 °C – 25 °C ±0.5 °C
• Dispensing onto the tube inner wall
• Operating mode: PIP
• Speed level: 5
Model Test tip
epT.I.P.S.
Testing
volume
Error of measurement
Systematic error Random error
± %± μL ± %± μL
0.5 µL – 10 µL 0.1 µL – 20 µL
medium gray
40 mm
1 µL 5.0 0.05 3.0 0.03
5 µL 3.0 0.15 1.5 0.075
10 µL 2.0 0.2 0.8 0.08
5 µL – 100 µL 2 µL – 200 µL
yellow
53 mm
10 µL 2.0 0.2 2.0 0.2
50 µL 1.0 0.5 0.8 0.4
100 µL 0.8 0.8 0.25 0.25
15 µL – 300 µL 15 µL – 300 µL
orange
55 mm
30 µL 2.5 0.75 1.0 0.3
150 µL 1.0 1.5 0.5 0.75
300 µL 0.6 1.8 0.25 0.75
50 µL – 1200 µL 50 µL – 1250 µL
green
76 mm
120 µL 6.0 7.2 0.9 1.08
600 µL 2.7 16.2 0.4 2.4
1200 µL 1.2 14.4 0.3 3.6