Operation
Eppendorf Xplorer
®
/Eppendorf Xplorer
®
plus
English (EN)
39
If you only hold the rocker pressed down during the blow-out (Blow) after
dispensing, the piston remains in the lowest position. The piston will only return
to the basic position after the rocker has been released. This presents the
following advantages:
If you pipette liquid that drains slowly, you can allow this liquid to flow out.
You can immerse the pipette tip in the target liquid to prevent unintentional
aspiration.
You can trigger the blow-out (Blow) any time the piston is in the basic
position. Press the rocker down.
Press the rocker in the opposite direction if you would like to stop aspiration
or dispensing while it is in progress. The display shows the volume present in
the pipette tip after the respective operation has been stopped. You can then
continue with liquid aspiration or dispensing.
Dispense the liquid against the tube inner wall if you are determining the
systematic and random errors.