54
Operation
Eppendorf Xplorer
®
/Eppendorf Xplorer
®
plus
English (EN)
Other volume limits may apply when using filter tips from other manufacturers. The
volume limit differs depending on the tip geometry and the position of the filter in the tip.
The volume limit is dependent on the additional aspirated volumes of the pipette. On the
Xplorer the individual nominal volumes result in different additional volumes.
ep Dualfilter T.I.P.S. Volume limit
Ads/Dis Reverse pipetting
10 µL – –
100 µL – –
200 µL – –
300 µL 270 µL 260 µL
1000 µL – 950 µL
1250 µL Long* – 1130 µL
5 mL 4.9 mL 4.6 mL**
5 mL Long* – –
10 mL* 9.75 mL 8.5 mL**
* When using these filter tips we recommend to use the adjustment epTIPS long in order
to increase the accuracy of dispensing. The volume limit stated applies only to the
changed adjustment.
** We recommend this volume limit even when performing reverse pipetting and using
the 5 mL and 10 mL epT.I.P.S. without a filter.
Xplorer
Single-channel and
multi-channel
Additional aspiration volumes
Ads/Dis Reverse pipetting
10 µL Approx. 0.6 µL Approx. 2 µL
100 µL Approx. 7 µL Approx .20 µL
300 µL Approx .20 µL Approx. 60 µL
1000 µL Approx. 65 µL Approx. 200 µL
1200 µL Approx. 140 µL Approx. 220 µL
5 mL Approx. 0.3 mL Approx. 1 mL
10 mL Approx. 0.55 mL Approx. 1.8 mL