14.1 Self Test of the Analyzer
Initiating a self test will test the key functions and
components of the analyzer. The process takes
approximately one minute and gives evaluation of
each parameter. The acceptable ranges of each
tested parameter or function can be found in the
following table.
Clicking or tapping any tested parameter’s small box
gives detailed information of the parameter, limits
and comments.
Figure 86. Diagnostics
By using the “Load last self-test” function you can check the results of the previous self-test
sequence.
Self-test checks parameters for the ranges defined below:
Big puffer time
Generate 3000 - 13000sec
Release 3000 - 5500sec
Small puffer time
Generate1 800 - 4000sec
Generate2 800 - 4000sec
Release 500 - 2000sec
Big puffer drift
Maximum 540 - 560mBar
Minimum 530 - 560mBar
Drift -5 - 15mBar
Small puffer drift
Maximum 225 - 235mBar
Minimum 215 - 235mBar
Drift -5 - 15mBar
Max vacuum P1
Maximum: 550 - 1000 mBar
Time: 0 – 6000 ms
Max vacuum P2
Maximum: 550 - 1000 mBar:
Time: 0 – 6000 ms
Pump status
Pump1 1 - 1
Pump2 1 - 1
Null pressures
Sheath -20 - 20mBar
Capillary -20 - 20mBar