Page 145 of 159
Mean Platelet Volume – MPV
(fl)
Average volume of individual platelets derived from
the PLT histogram
Thrombocrit – PCT
(percentage, absolute)
Calculated from the PLT and MPV values
PCTpercentage = PLT x MPV x 100
PCTabsolute = PLT x MPV
White blood cell, 4 part differential:
LYM (%) : lymphocytes
MON (%) : monocytes and some
eosinophils
NEU (%) : neutrophil granulocytes
EOS(%):eosinophil granulocytes
Relative value are calculated from the optical 4 diff
scatter populations
BAS (%):basophil granulocytes
Relative value is calculated from the basofil optical
scatter populations
20.4.2 Technical Data
20.4.2.1 Main Unit
Sample volume Closed-, and Open-mode: 110 l
Sample small module – only open mode: 25 µl
Sample type Human whole blood (K
3
-EDTA anticoagulant)
Tube Identification By means of the front panel keyboard (enter ID)
By means of the barcode labels (manual and/or auto-sampler)
Sampling method Ceramic shear valve with 3 separated primary loops
Measured parameters CBC+5DIFF mode (24 parameters):
WBC, LYM, MON, NEU, EOS, BAS, LYM%, MON%, NEU%,
EOS%, BAS%
RBC, HCT, MCV, RDWsd, RDWcv HGB, MCH, MCHC
PLT, PCT, MPV, PDWsd, PDWcv
Throughput 60 tests/hour
Measurement method Volumetric impedance change for WBC, RBC, PLT
Spectrophotometry for HGB
Light scattering 4-diff measurement: LYM, MON, NEU, EOS
Light scattering BASO measurement
Aperture diameter WBC: 80
m, RBC/PLT: 70 m