EasyManua.ls Logo

ERBA ELite 5 - Approved Reagents and Control Materials

ERBA ELite 5
165 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
ELite 5, User Manual
Page 22 of 159
Par
ameter
Symbol
Measurement method
Red Blood Cell count RBC Impedance measurement
Hemoglobin HGB Photometric measurement
Hematocrit HCT Calculated
Mean Corpuscular Volume MCV Derived
Mean Corpuscular Hemoglobin MCH Calculated
Mean Corpuscular Hemoglobin
Concentration
MCHC Calculated
Red Blood cells Distribution Width RDWcv Derived
Red Blood cells Distribution Width RDWsd Derived
Platelet count PLT Impedance measurement
Mean Platelet Volume MPV Derived
Platelet Distribution Width**
(for Research Use Only)
PDWcv Derived
Platelet Distribution Width**
(for Research Use Only)
PDWsd Derived
Thrombocrit**
(for Research Use Only)
PCT Calculated
Large platelet count PLCC PLT count above 12 fl in
size
Large platelet percentage PLCR Calculated
**The PDWcv, PDWsd, and PCT parameters are considered to be for Research
Use Only (RUO), and will be marked on the screen, printout, database, and LIS
transmission.
Table 1. ‘ELite 5’ Analyzer Parameters
4.1.2 Approved Reagents and Control Materials
Always use reagents, control and calibration materials that are recommended and approved
by the manufacturer. The ‘ELite 5’ analyzer, controls, calibrator and reagents are part of a
system and are carefully designed and selected for optimal performance. Using unapproved
reagents or controls may cause false flagging or incorrect, inaccurate results.
Only original Erba Lachema reagents may be used in the ‘ELite 5’ analyzer. For additional
details about these materials, please contact the manufacturer directly.
4.1.3 Original Erba Lachema Reagents
Use only the original Erba Lachema reagents listed below to ensure proper
performance of the ‘ELite 5’ analyzer. Reagents are manufactured and provided by
Erba Lachema s.r.o., and are for in vitro diagnostic use only. For using leave the
reagent at room temperature for at least 12 hours. Do not use reagent once frozen.
Kat. číslo
Název
Objem
HEM00006
Erba Lyse-5P 5 l
HEM00002
Erba Diff-5P 1 l
HEM00005
Erba Hypoclean CC 100 ml

Table of Contents

Related product manuals