4.2.8 Front Assembly
Shear Valve
Aspiration Needle
Sample Rotor
Figure 15. Front Assembly Behind the Front Panel
Syringe pumps control the volume and flow rate of reagents through the pneumatic
components.
Shear valve is an integrated multi functional valve for segmenting precise volumes of
blood.
4.3 Small Sample Mode (SSM) – optional
Instruments which have the small sample module (SSM) installed can be used for taking
sample from a small amount of whole blood sample, which is not enough for normal
sampling.
SSM measurements are only open vial measurements.
At least 250 µl of sample is necessary for reliable sampling.
The analyzer will aspirate 25 µl of whole blood and put it in to the pre-dilution
chamber to dilute it to 1:6.
After pre-dilution, the needle will take 100 µl of diluted sample into the shear valve
and ´ELite 5´ is going to proceed with the normal measuring process.