Printer: Selects which printer to print to
Printer status: Display the currently selected printer status
Color printing: Selects whether to print in color
Double sided printing: Selects whether to print multi-page printouts on both sides of a page
Items in queue: Number of items in the printer queue
Cancel all jobs: Cancels all items in the selected printer’s queue
Pre-printed paper´s header length (px): Reserves space (pixels) at top of page for pre-printed
letterhead
Automatic print: Selects whether to print for every new sample processed
Logo visible: Selects whether to print or omit graphical logo to printout
Warning flags are visible: Selects whether to print or omit warning flags to printout
Diagnostic flags are visible: Selects whether to print or omit diagnostic flags to printout
Refresh printers list: Refreshes the list of displayed printers
Profile limits: Enter normal ranges for various profile limits
X-B: Change X-B limits and targets
User: Add and manage users of the ‘ELite 5’ analyzer (only available to administrator users)
Delete user: Initiates deletion of a user account
Administrator: Selects or unselects administrator privileges for the currently selected user
See section 13 for more details about ‘ELite 5’ analyzer settings
Maintenance
Cleaning: Perform extended cleaning of various pneumatic components
Draining: Drain various internal system reservoirs
Empty chamber: Empty RBC, WBC, and mix chambers
Prime: Prime one or all reagent reservoirs
Fill: Fills the fluidic components of the system with reagents
Show MS On-Screen Keyboard: Brings up a vertical keyboard
Touch screen: Calibrate the touch screen location for taps and button presses
Reset Auto Sampler: Sets Auto Sampler to home position
See section 15 for more details about ‘ELite 5’ maintenance procedures
Calibration
Calibrate
Calibration mode: Select type of calibration procedure to run
Calibration type: Select human blood or calibration material
Target values: Enter the target values for each calibrated parameter
Cancel: Cancel a calibration procedure
Next: Initiates processing of calibration run
View Calibrations: View and delete previous calibrations
See section 10 for more details about ‘ELite 5’ calibration
QC
QC Measure: Initiate processing of a control sample
QC Reference select: Select which stored QC reference this measurement belongs to
Set QC Reference: Create a new stored QC reference
Load QC reference: load QC reference values from removable media such as a flash drive
Save reference: Save entered QC reference values
View QC references: Browse stored QC references
Delete: Delete a stored QC reference
Details: View detailed results for a stored QC reference
View QC data: Browse individual QC sample measurements
Delete: Delete currently selected QC run
View QC diagrams: View QC diagrams for currently selected control
Details: View detailed result screen of a QC run
View QC diagrams: View Levy-Jennings diagrams of QC data for the currently selected QC
reference
View QC data: Browse individual QC sample measurements
Next diagrams: Browse the next set of QC diagrams