EasyManua.ls Logo

ERBA ELite 5 - Page 50

ERBA ELite 5
165 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
ELite 5, User Manual
Page 50 of 159
Figure 312. Prepare for shipment
If the ‘ELite 5’ analyzer is powered off without a proper shutdown or prepare
for shipment procedure, the following may occur:
The pneumatic components may lock in an undefined state. The next
startup and pneumatic initialization may take longer with increased
reagent consumption.
The ‘ELite 5’ application software and/or the Windows
®
XP
®
Embedded operating system can suffer damage or data loss.
The ‘ELite 5’ analyzer may have to be restarted by a service engineer
after an improper shut down or power off sequence.
The ‘Prepare for Shipment’ procedure should be used for following reasons:
The reagents may spill out if the ‘ELite 5’ is tilted, potentially causing electrical short
circuit or corrosion.
Reagents can flow out of the ‘ELite 5’ if the reagent intakes are disconnected.
Evaporation may cause reagents left in the pneumatic components in the ‘ELite 5’
analyzer to become concentrated and form crystals that increase blank results,
damage pneumatic components, and may require a service engineer to restore
proper operation.

Table of Contents

Related product manuals