EasyManua.ls Logo

ERBA ELite 5 - Page 81

ERBA ELite 5
165 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
ELite 5, User Manual
Page 81 of 159
Warning flags
Warning flags are raised when the analyzer detects
a problem in the analysis process that may cause
incorrect or suspect results or the parameters in the valid blank measurement was too high.
Flag
Meaning
Cause
Action
l PLT URI
The PLT-RBC gap is not clearly detectable on the
PLT-RBC histogram: small MCV, fractured RBC-s,
aggregated PLT-s (cold blood), side-effect of blood-
transfusion.
Check the sample quality. Repeat the sample. If the
problem persists then perform a manual count on a
stained smear.
M
Out of RBC linearity
range
The RBC coincidence is over the limit: too high RBC
count.
Check the sample homogeneity. Repeat the sample
with manual pre-dilution.
m
Close to RBC linearity
range
The RBC coincidence is close to the limit: too high
RBC count.
Treat RBC result as ‘reduced reliability’. Check the
sample homogeneity. Repeat the sample with manual
pre-dilution.
N High angle noise
The flow in the optical head is not clean or contains
bubbles
Perform a blank measurement. If noise still appears,
make a flow cell backwash.
n Low angle noise
The flow in the optical head is not clean or contains
bubbles
Perform a blank measurement. If noise still appears,
make a flow cell backwash.
O
High linearity range
limit exceeded
If any of RBC,PLT or WBC results are higher than
high linearity range limits.
Perform manual pre-dilution of sample and re-
measure it.
o
Low linearity range
limit exceeded
If any of RBC,PLT or WBC results are lower than Low
linearity range limits.
None
p PLT blank high Last accepted blank result: PLT 15 * 103 cellsl
Check cleanliness of the reagents and the ‘ELite 5’.
Perform cleaning of the ’ELite 5’. Repeat blank
measurement. See chapter: 6.2.5:Pneumatics start
and blank measurement
Treat PLT result as ‘reduced reliability’.
Q MON - LYM alert
The MON and LYM populations are not clearly
distinguishable on the scatter-gram.
If other parameters predict a ‘normal’ sample then
perform a cleaning procedure and repeat the
measurement.
If MON, MON%, LYM, LYM% results reported then
treat them as ‘reduced reliability’. If not all the
parameters reported or the sample is predicted as
pathological then perform a manual count on a stained
smear.

Table of Contents

Related product manuals