ELite 5, User Manual
Page 98 of 159
Figure 71. Set QC Reference Panel
The ‘QC name’ is the name that will be displayed for
selecting the control to associate with a QC
measurement. Enter the identifying information, and the
target values and tolerance ranges for any parameters
that are to be controlled with the ‘ELite 5’ QC system. To
exclude a parameter from the QC procedure, leave its
target value and tolerance range field empty. Press the
‘Save reference’ button after data entry is completed.
Tapping the “Load QC reference” button will allow filling
QC material data electronically – using a file provided by
the manufacturer of the analyzer. A dialog will open.
Browse to the correct folder, and select the file by
tapping the check box in front of the filename.
The file contains information for all three levels of the
control material (Low, Normal and High). All three levels
will be loaded into the system and will become
selectable in the QC menu.