Measurement
28 MMS
®
Inspection DPM High
6Measurement
The gage model MMS Inspection DPM High is for continuously measuring tem-
perature (Ta, Ts) and humidity (RH) to determine the dew point (Td) and the dif-
ference between surface temperature and dew point (Ts-Td). You can log all
these climate values over a prolonged period, monitor specification limits and
emit an optical warning if limits are exceeded (alarm function).
The following factors are measured: air temperature (Ta), relative humidity (RH)
and surface temperature (Ts)
From the measured factors, the dew point (Td) and the difference between sur-
face temperature and dew point (Ts-Td) are calculated.
All measured and calculated climate factors are saved simultaneously (manu-
ally or continuously).
Notes on measuring
Measurement errors
Measurement errors occur if ambient air
cannot flow through the housing grille in the
lower area of the gage!
Keep the housing grille in the lower area of
the gage clear. Avoid covering with hands,
material or other objects.
Allow ambient air to flow unhindered through
the housing grille in the lower area of the
gage.
Observe the instructions for handling the
gage during a measurement, described in
Chapter "Correctly holding the gage during
the measurement" on page 30
Values outside the specification limits are dis-
played in red.
Limit violations are indicated by a red flashing signal LED, if the alarm
function is activated.
The probes for air tempera-
ture and relative humidity are
protected by the housing
grille