10-14
Multimeter disturbance:
VDC, VAC, and VAC+DC with STL120 and short ground lead.
OHM, CONT, DIODE, and CAP with STL120, and black test lead to COM.
Table 3
Disturbance less than 1% of full scale E= 3 V/m E= 10 V/m
Frequency range 10 kHz to 27 MHz
VDC, VAC, VAC+DC
OHM, CONT, DIODE
CAP
500 mV to 1250V
50 Ω to 30 MΩ
50 nF to 500 μF
500 mV to 1250V
50 Ω to 30 MΩ
50 nF to 500 μF
Frequency range 27 MHz to 1 GHz
VDC, VAC, VAC+DC
OHM, CONT, DIODE
CAP
500 mV to 1250V
50 Ω to 30 MΩ
50 nF to 500 μF
500 mV to 1250V
50 Ω to 30 MΩ
50 nF to 500 μF
Disturbance less than 1% of full scale E= 3 V/m E= 1 V/m
Frequency range 1.4 GHz to 2 GHz
VDC, VAC, VAC+DC
OHM, CONT, DIODE
CAP
500 mV to 1250V
50 Ω to 30 MΩ
50 nF to 500 μF
Frequency range 2 GHz to 2.7 GHz
VDC, VAC, VAC+DC
OHM, CONT, DIODE
CAP
500 mV to 1250V
50 Ω to 30 MΩ
50 nF to 500 μF
Test tool ranges not specified in table 3 may have a disturbance of more than 10% of full scale.