EasyManua.ls Logo

Fluke 125 - Making Measurements; Connecting the Inputs

Fluke 125
128 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Fluke 125
Users Manual
2-4
Making Measurements
The reading area displays the numeric readings of the
chosen measurements on the waveform that is applied to
the input jack.
Connecting the Inputs
Voltage measurements
See figure 2-3. For proper grounding connect the short
ground leads
d
to the same ground potential. You can
also use test lead
c
for grounding. See also Using
Proper Grounding in chapter 9.
Ohm (Ω), continuity, diode, and capacitance
measurements
See figure 2-4. Use the red shielded test lead from input A
and the black unshielded ground lead from COM
(common).
Current measurements
See figure 2-5. Select the probe setting that matches the
used current clamp and its setting (e.g. 1 mV/A), see
Chapter 1, Measurement Probes and Settings.
Temperature measurements
See figure 2-6. Use a 1 mV/°C temperature transmitter
(not available in all countries) to get the correct
temperature reading.
Power measurements
See figure 2-7. Select correct probe settings for voltage
measurement on input A, and current measurement on
input B.
1
2
2
Figure 2-3. Volts Measurement Setup

Table of Contents

Other manuals for Fluke 125

Related product manuals