EasyManua.ls Logo

Fluke 125 - Ethernet Twisted Pair;10 Baset

Fluke 125
128 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Fluke 125
Users Manual
A-10
COAX
COAX
COAX
PM9083
BB120
PASSIVE
TAP
PASSIVE
TAP
DEVICE 1 DEVICE 2
Ethernet Twisted pair/10BaseT
To check this bustype, ScopeMeter Channel A is on and
DC coupled. The recommended probe is VPS40 (10:1),
Channel A is on and DC coupled. The probe tip must be
connected to the TD+ or RD+ wire; the COM input (or
Black Ground Lead) to the TD– or RD– wire. Data traffic
is not always continuous.
TD+
RD+
TD-
RD-
OR
CABLE
The wires can be reached with Back Probe Pins TP88
(optional) at screw terminals at a device’s wire entry
point: for instance at a Junction Box. The right-hand
figure shows pinning and wire colors of a RJ-45
connector.
Pin 1: TD+ (brown)
Pin 2: TD- (brown/white)
Pin 3: RD+ (orange)
Pin 4
Pin 5
Pin 6: RD- (orange/white)
Pin 7
Pin 8