Maintaining the Test Tool
Using and Adjusting 10:1 Scope Probes
8
8-5
Note
Ensure that the battery pack is placed in the
battery compartment as shown in Figure 8-2.
8. Reinstall the battery cover and secure the screw.
Using and Adjusting 10:1 Scope Probes
Note
The 10:1 voltage probe VPS40 that is supplied
with Fluke 125 is always adjusted correctly to the
Test Tool and needs no further adjustment.
You need to adjust other 10:1 scope probes for optimal
response.
Warning
To avoid electrical shock use the BB120
Banana-to-BNC adapter (delivered with the
test tool) to connect a 10:1 scope probe to the
input of the test tool.
To adjust probes, do the following:
Connect the 10:1 scope probe from the gray input B
jack to the red input A jack. Use the red 4-mm banana
adapter (delivered with the probe) and the banana-to-
BNC adapter (BB120). See Figure 8-3.
BB120
Figure 8-3. Adjusting Scope Probes