Fluke 125
Users Manual
10-4
Advanced Scope Functions
Display Modes
Normal ....... Captures up to 40 ns glitches and displays
analog-like persistence waveform.
Smooth ...... Suppresses noise from a waveform.
Envelope.... Records and displays the minimum and
maximum of waveforms over time.
Auto Set (Connect-and-View)
Continuous fully automatic adjustment of amplitude, time
base, trigger levels, trigger gap, and hold-off. Manual
override by user adjustment of amplitude, time base, or
trigger level.
Dual Input Meter
The accuracy of all measurements is within ± (% of
reading + number of counts) from 18 °C to 28 °C.
Add 0.1x (specific accuracy) for each °C below 18 °C or
above 28 °C. For voltage measurements with 10:1 probe,
add probe uncertainty +1%. More than one waveform
period must be visible on the sceen.
Input A and Input B
DC Voltage (VDC)
Ranges.........................500 mV, 5V, 50V, 500V, 1250V
Accuracy .......................................... ±(0.5% +5 counts)
Normal Mode Rejection (SMR)...........................>60 dB
@ 50 or 60 Hz ±1%
Common Mode Rejection (CMRR)........ >100 dB @ DC
>60 dB @ 50, 60, or 400 Hz
Full Scale Reading .....................................5000 counts
True RMS Voltages (VAC and VAC+DC)
Ranges.........................500 mV, 5V, 50V, 500V, 1250V
Accuracy for 5 to 100% of range
DC coupled:
DC to 60 Hz (VAC+DC) ............... ±(1% +10 counts)
1 Hz to 60 Hz (VAC) .................... ±(1% +10 counts)