Fluke 125
Users Manual
2-24
Making Rise Time Measurements
To measure rise time, do the following:
c
From Scope mode, display the
Cursor Key functions.
d
Press to highlight (rise time).
Observe that two horizontal
cursors are displayed.
e
If only one trace is displayed,
select MANUAL or AUTO. AUTO
automatically does steps 4 to 6.
For multiple traces select the
required trace A or B.
f
Move the upper cursor to 100% of
the trace height. A marker is
shown at 90%.
g
Highlight the other cursor.
h
Move the lower cursor to 0% of
the trace height. A marker is
shown at 10%.
Figure 2-20. Risetime Measurement with Cursors
The reading now shows the risetime from 10%-90% of the
trace amplitude and the voltage at the cursors in relation to
the zero icon (-). See Figure 2-20.
i
Turn off the cursors.