Remote Operation
Supported SCPI Commands 5
5-33
b Test current is dc with negative polarity (DCN)
NO is returned when the generated test signal is not valid.
[SOUR]:SAF:LIN:CURR?
Description: Returns the measured maximum value of peak current flowing
through the selected resistance. If not already selected, it also switches
the Calibrator to the Line Impedance function.
Query: SAF:LIN:CURR?
Note: The maximum value can be cleared with the SAF:LINE:CLE
command.
[SOUR]:SAF:LIN:CLE
Description: This command clears the maximum value of peak current. If not
already selected, it also switches the Calibrator to the Line Impedance
function.
Example: SAF:LIN:CLE
[SOUR]:SAF:LIN[:CURR]:PFC?
Description: Returns the value of prospective fault current. If not already selected,
it also switches the Calibrator to the Line Impedance function.
Query: SAF:LIN:PFC?
[SOUR]:SAF:LIN:MAN(?) <DNPD>
Description: This command sets the manual value of line residual impedance. If not
already selected, it also switches the Calibrator to the Line Impedance
function.
Parameters: <DNPD> The value of residual impedance in ohms.
Example: SAF:LIN:MAN 0.87 Set manual compensation value to 0.87 Ω.
Query: SAF:LIN:MAN? Returns the manual compensation value.
[SOUR]:SAF:LIN:SCAN?
Description: Returns the scanned line residual impedance value. If not already
selected, it also switches the Calibrator to the Line Impedance
function.
Query: SAF:LIN:SCAN?
[SOUR]:SAF:LOOP[:LEV](?) [<DNPD>]
Description: This command sets the resistance value as specified by DNPD for the
Loop Impedance function. If not already selected, it also switches the
Calibrator to the Loop Impedance function.
Parameters: <DNPD> The value of resistance in ohms. The
Calibrator selects the closest resistance
value.
Example: SAF:LOOP 0.1 Set the Loop Impedance function with
100 mΩ.