Calibrating Instruments
Measuring with the Built-in Multimeter 4
4-31
Cursor icons appear above and below one of the digits in the frequency display.
2. Use the cursor keys, rotary knob or keypad to set the frequency value.
3. Press the Exit softkey to exit the frequency edit mode.
To switch between grounded and ungrounded (floating) output:
1. Press the Setup softkey.
2. Using the cursor keys or rotary knob, highlight Calibrator and either press the
Select softkey or press in on the rotary knob.
3. Using the cursor keys or rotary knob, move the cursor to Output GND and either
press the Select softkey or press in on the rotary knob.
4. Using the cursor keys or rotary knob, move the cursor to GND On or GND Off and
either press the Select softkey or press in on the rotary knob.
5. Return to the main display by pressing the Exit softkey repeatedly.
Measuring with the Built-in Multimeter
The Calibrator is equipped with an ac/dc voltage, current, current leakage, and time
interval meter. The meter has three selectable modes as well as a help guide:
Multimeter: The meter will measure up to 1100 volts rms and up to 30 amps rms
simultaneously. This makes it possible to measure power consumption and line
current simultaneously when calibrating Appliance testers.
HIPOT LC: The meter will simultaneously measure output voltage and leakage
current of hipot testers up to 300 mA.
HIPOT Timer: The meter measures the time interval of hipot testers up to 999
seconds.
Help: The help guide offers basic information about the selected function.
The meter’s input terminals and function selection button are located in the lower left
corner of the front panel (See items 4 and 6 in Table 3-1 of Chapter 3).
Note
When measuring with the built-in multimeter, the meter terminals are
always connected. Measurements are taken whether the Calibrator is in
standby or operate mode.
Function Selection
To set the meter mode:
1. Press the Mode softkey to display the Available Modes menu (see Figure 4-20).
2. Using the cursor keys or rotary knob, highlight the mode.
3. Press the Select softkey or press in on the rotary knob to select the highlighted
mode.

Table of Contents

Other manuals for Fluke 5320A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke 5320A and is the answer not in the manual?

Fluke 5320A Specifications

General IconGeneral
BrandFluke
Model5320A
CategoryTest Equipment
LanguageEnglish

Related product manuals