5320A
Users Manual
3-14
Setting the Beeper Volume
To set the beeper volume:
1. After entering the setup menu, press P or Q or rotate the rotary knob to highlight
“Beeper Volume” in the list of setup selections.
2. Press the Select softkey or press in on the rotary knob.
3. Set the beeper’s volume between 0 and 15 using the rotary knob, the cursor keys or
by entering the value directly with the keypad.
Note
For safety warnings and messages, the beeper will still sound even when
the beeper volume is set to zero.
4. Press the WRITE softkey to set the beeper volume and return to the setup menu.
Setting the Display Brightness
To set the brightness of the display:
1. After entering the setup menu, press P or Q or rotate the rotary knob to highlight
“Display Brightness” in the list of setup selections.
2. Press the Select softkey or press in on the rotary knob.
3. Set the display’s brightness between 0 and 7 using the rotary knob, the cursor keys,
or by entering the value directly with the keypad.
4. Press the Write softkey to set the display’s brightness and return to the setup menu.
Setting the Calibration Password
The calibration password is set to “00000” when shipped from the factory. Only when the
password is changed to a non-zero value must a password be entered to access the
calibration mode. The purpose of the password is to prevent unauthorized users from
changing the calibration settings.
To set the calibration code:
1. After entering the setup menu, press P or Q or rotate the rotary knob to highlight
“Calibration Password” in the list of setup selections.
2. Press the Select softkey or press in on the rotary knob.
3. If the calibration password is set to a non-zero value, you will be prompted to enter
the present password.
4. Use the keyboard to enter a new code and press E. Make sure what is displayed
in the display and what you think you entered are the same. After completing the next
step, you must have the correct password to gain access to the calibration functions.
5. Press the Write softkey to set the calibration password and return to the setup menu.
To exit setting the password without changing it, press Exit.
Viewing the Device Information
To view Calibrator information (serial number, software versions, and installed options):
1.
After entering the setup menu, press P or Q or rota
te the rotary
knob to highlight
“Device Information” in the list of setup selections.
2. Press the Select softkey or press in on the rotary knob.
3. After viewing the information, press the Exit softkey.

Table of Contents

Other manuals for Fluke 5320A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke 5320A and is the answer not in the manual?

Fluke 5320A Specifications

General IconGeneral
BrandFluke
Model5320A
CategoryTest Equipment
LanguageEnglish

Related product manuals