Remote Operation
Supported SCPI Commands 5
5-39
Query: SAF:HIPT:VOLT?
[SOUR]:SAF:HIPT:[VOLT]:MAX?
Description: Returns the maximum peak measured value of voltage using standard
exponential format. For example, 230 V is returned as
2.300000e+002. Use the SAF:HIPT:CLEar command to clear the
maximum value. If not already selected, it also switches the Calibrator
to the HIPOT timer function.
Query: SAF:HIPT:MAX?
[SOUR]:SAF:HIPT:CLE?
Description: Clears the measured time and maximum peak value of measured
voltage. If not already selected, it also switches the Calibrator to the
HIPOT timer function.
Example: SAF:HIPT:CLE
Using SYSTem Commands
The SYSTEM commands are used to read and control miscellaneous Calibrator
functions.
SYSTem
DATE (?) <DNPD>,<DNPD>,<DNPD>
TIME (?) <DNPD>,<DNPD>,<DNPD>
ERRor?
REMote
RWLock
LOCal
SYST:DATE(?) <DNPD>,<DNPD>,<DNPD>
Description: This command sets the system date of the Calibrator.
Parameters: <DNPD> Year as YYYY
<DNPD> Month as MM
<DNPD> Day as DD.
Example: SYSR:DATE 2006,11,05 Sets the Calibrator’s date function to
November 11, 2006.
Query: SAF:DATE? Returns the date in the Calibrator.
SYST:TIME(?) <DNPD>,<DNPD>,<DNPD>
Description: This command sets the system time of the Calibrator.
Parameters: <DNPD> Hour as HH (00 to 23)
<DNPD> Minutes as MM (00 to 59)
<DNPD> Seconds as SS (00 to 59).
Example: SYSR:TIME 15,36,14 Sets the Calibrator’s time function to 15
hours, 36 minutes and 14 seconds (3:36:14
PM).