Operator Maintenance
Verifying Calibrator Operation 6
6-13
Tables of Limits
The following tables are used to verify the operation of the Calibrator.
Table 6-4. Low Resistance Source Limits
Nominal value
Required standard
ohmmeter uncertainty
Lower limit Upper limit
0.100 Ω
0.02 %
89.7 mΩ 110.3 mΩ
0.200 Ω
0.02 %
189.4 mΩ 210.6 mΩ
0.400 Ω
0.02 %
388.8 mΩ 411.2 mΩ
0.800 Ω
0.02 %
787.6 mΩ 812.4 mΩ
2.000 Ω
0.02 %
1.984 Ω 2.016 Ω
4.000 Ω
0.02 %
3.978 Ω 4.022 Ω
8.000 Ω
0.02 %
7.974 Ω 8.026 Ω
10.00 Ω
0.02 %
9.970 Ω 10.030 Ω
20.00 Ω
0.02 %
19.95 Ω 20.05 Ω
40.00 Ω
0.02 %
39.91 Ω 40.09 Ω
80.00 Ω
0.02 %
79.83 Ω 80.17 Ω
100.0 Ω
0.02 %
99.79 Ω 100.21 Ω
200.0 Ω
0.02 %
199.6 Ω 200.4 Ω
400.0 Ω
0.02 %
399.2 Ω 400.8 Ω
800.0 Ω
0.02 %
798.4 Ω 801.6 Ω
1000 Ω
0.02 %
998.0 Ω 1002.0 Ω
2000 Ω
0.02 %
1996 Ω 2004 Ω
4000 Ω
0.02 %
3992 Ω 4008 Ω
8000 Ω
0.02 %
7984 Ω 8016 Ω
10000 Ω
0.02 %
9980 Ω 10020 Ω