Query: SAF:TIME? Returns the time in the Calibrator.
SYST:ERR?
Description: This command queries the Calibrator’s error queue to retrieve the first
error in the queue. The error queue stores errors as first-in first-out
(FIFO). The error is deleted after being read. In the case of error queue
overflow, the last error in the queue is –350,”Queue overflow”. The
earliest errors remain in the queue, while the most recent error is
discarded.
Query: SYST:ERR? Returns first error in the error queue as an
error code followed by an error message.
SYST:REM
Description: This command sets the Calibrator in remote mode for remote control
through the RS-232 or Ethernet ports. All keys on the front panel
except the LOCAL key are disabled.
Note
It is not possible to send or receive data over the RS-232 or Ethernet
ports when the calibrator is not in remote mode.
SYST:RWL
Description: This command sets the Calibrator in remote mode for remote control
through the RS-232 or Ethernet ports. All keys on the Calibrator’s
front panel except the LOCAL key, are disabled.
SYST:LOC
Description: This command returns the calibrator to the local mode. This command
is used only when controlling the Calibrator through the RS-232 or
Ethernet ports.
Using the STATus Subsystem
This subsystem is used to enable bits in the Operation and Questionable Event
registers. The Operation and Questionable Event, Enable and Condition registers can
be interrogated to determine their state. Table 5-6 give a summary of the Calibrator’s
STATUS commands.
STAT:OPER:EVEN?
Description: This query command returns the content of Operational Data Event
register. It is a decimal value which corresponds to the binary-
weighted sum of all bits set in the register. Register is cleared after
this query.
STAT:OPER:ENAB? <DNPD>
Description: This command enables bits in the Operational Data Enable register.
Selected bits are summarized at bit 7 (OSS) of the IEEE 488.2 Status
Byte register.
Parameters: <DNPD> Represents the register bits as a decimal
number.

Table of Contents

Other manuals for Fluke 5320A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke 5320A and is the answer not in the manual?

Fluke 5320A Specifications

General IconGeneral
BrandFluke
Model5320A
CategoryTest Equipment
LanguageEnglish

Related product manuals