5320A
Users Manual
5-42
The reply is formatted as follows:
FLUKE,5320A,100002,1.12+1.01+1.01+1.02
*OPC
Description: This command sets the OPC bit in the ESR (Event Status Register)
when all pending operations are complete.
*OPC?
Description: This command returns “1” to the output queue after all pending
operations are complete.
*WAI
Description: This command prevents the Calibrator from executing any further
commands or queries until all previous remote commands have been
executed.
*RST
Description: This command resets the Calibrator to its initial state.
*TST?
Description: This command launches an internal self-test and returns the self-test
results (“0” = pass or “1” = fail).
*STB?
Description: This query command returns a number representing the STB register
contents. The STB register carries the MSS bit status.
*SRE <value>
Description: This command sets the Service Request Enable register. Since bit 6 is
not used, the maximum value is 191.
*SRE?
Description: This query command returns the Service Request Enable Register
number.
*ESR?
Description: This query command returns the contents of the Event Status Register
and clears the register.
*ESE <value>
Description: This command programs the Event Status Enable register bits.
Parameter value is a number in the range of 0 – 255.
*ESE?
Description: This query command returns the value in the Event Status Enable
register.

Table of Contents

Other manuals for Fluke 5320A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke 5320A and is the answer not in the manual?

Fluke 5320A Specifications

General IconGeneral
BrandFluke
Model5320A
CategoryTest Equipment
LanguageEnglish

Related product manuals