5320A
Users Manual
4-34
ehq061.bmp
Figure 4-22. Multimeter Display with Hipot Leakage Current Measurement
4. Connect the test leads to the V and COM jacks.
In hipot timer mode, the Calibrator measures the time interval in seconds. To make a
hipot timer measurement using the built-in multimeter:
1. Press M.
2. Select the HIPOT Timer mode.
3. If necessary, press the AC/DC softkey to select between ac or dc mode.
4. Connect the test leads to the V and COM jacks.
5. Set the hipot UUT to less than 1000V and select a timer mode. Activate the timer
mode on the UUT when ready.
ehq67.bmp
Figure 4-23. Multimeter Display with Hipot Timer Measurement
The current hipot output voltage and its maximum value during the measurement period
are displayed in the PARAMETERS area of the display. To clear the time interval
measurement, press the Clear softkey.