IRECTION 5604324, REVISION 11 DRAFT (JANUARY 24, 2019) SERVICE MANUAL
Chapter 4 - Functional Checks 4-40
9) Press the Reset button in Display tab -> Display options -> Details
10) Press the OK when pop-up the window like section 4-42
11) Press “Calibrate...” in Display tab -> Display options -> Details
Figure 4-41 Tablet PC Settings dialog
Figure 4-42 Digitizer Calibration Tool