GERAFT LOGIQ P9/P7
D
IRECTION 5604324, REVISION 11 DRAFT (JANUARY 24, 2019) SERVICE MANUAL
8-140 Section 8-26 - Replacement of 4D
2) Unscrew five screws to remove the DC4D from the IO BOX. Refer to following figure.
Figure 8-164 Screw points of DC4D Assy
3) Carefully lift up the DC4D unit.
8-26-4 Installation Procedure
Install components in reverse order of removal.
8-26-5 Functional Check
Refer to section 4-4 for Functional Diag Test List.