Loading...
GEDRAFT LOGIQ P9/P7
D
IRECTION 5604324, REVISION 11 DRAFT (JANUARY 24, 2019) SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-131
NOTE: There are various types of VPD EEPROM integrated in the system.
8-23-6-1 Accessing EEPROM
1) Exit to Desktop (Refer to section 5-10 for details)
2) Start VPD Editor by clicking icon on the desktop. (“Vpdedit”)
3) Select COMEXP from device grid.
Board Name Main VPD Contents Replaceable Physical VPD location
MCB FuncRev
No
Integrated with MCB
MPI FuncRev
No
Integrated with MPI
MDC FuncRev
No
Integrated with MDC
COMEXP FuncRev, , SSN (System Serial Number
Yes
On MCB board as replaceable
EEPROM.
MIO FuncRev
No
Integrated with MIO
DC4D FuncRev
No
Integrated with DC4D
OPIOR FuncRev
No
Integrated with OPIO

Table of Contents