GEDRAFT LOGIQ P9/P7
D
IRECTION 5604324, REVISION 11 DRAFT (JANUARY 24, 2019) SERVICE MANUAL
Chapter 2 - Site Preparation 2-5
2-2-4 Probe Environmental Requirements
2-2-5 Time and Manpower Requirements
Site preparation takes time. Begin Pre-installation checks as soon as possible. If possible, allow six
weeks before delivery, for enough time to make necessary changes.
Table 2-5 2D Probe Environmental Requirements
Operational
Temperature
10
o
- 35
o
C
for 2D probe
18
o
- 30
o
C
for 4D probe
Humidity
10 - 80% (non-condensing) for 2D probe
30 - 80% (non-condensing) for 4D probe
Pressure 700 - 1060hPa
NOTICE
!! NOTICE:
SYSTEMS AND ELECTRONIC PROBES ARE DESIGNED FOR STORAGE TEMPERATURES OF
0 TO + 55 degrees C. WHEN EXPOSED TO LARGE TEMPERATURE VARIATIONS, THE
PRODUCT SHOULD BE KEPT IN ROOM TEMPERATURE FOR 10 HOURS BEFORE USE.
CAUTION
!! CAUTION:
Have two people available to deliver and unpack the LOGIQ P9/P7 ultrasound system.
Attempts to move the unit considerable distances (or on an incline) by one person alone, could
result in personal injury and/or damage to the system.

Table of Contents

Other manuals for GE LOGIQ P7

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ P7 and is the answer not in the manual?

GE LOGIQ P7 Specifications

General IconGeneral
BrandGE
ModelLOGIQ P7
CategoryMedical Equipment
LanguageEnglish

Related product manuals