GERAFT LOGIQ P9/P7
D
IRECTION 5604324, REVISION 11 DRAFT (JANUARY 24, 2019) SERVICE MANUAL
7-12 Section 7-4 - Screen Captures and Logs
7-4-3-5 System Error
7-4-3-5-1 Cause
Unrecoverable error occurred and unable to continue scan. Error dialog does not indicate the root
cause of system error. To identify the root cause clearly, it is necessary to investigate the log files.
7-4-3-5-2 What to do
1) Press “OK”, turn off the console and reboot.
2) If same issue persists, then shut down the system, and turn off the system breaker.