GEDRAFT LOGIQ P9/P7
D
IRECTION 5604324, REVISION 11 DRAFT (JANUARY 24, 2019) SERVICE MANUAL
Chapter 5 - Components and Functions (Theory) 5-53
This table shows all the elements available on FRU Status with descriptions.
For more information, see:
Voltage on page 5-52
Home on page 5-42
Table 5-19 FRU Status
Element DESCRIPTION
Name Name of the FRU.
Status
Status of the voltage for the FRU. Valid values are:
• Normal – Indicates that the voltage is within the allowable range.
• Warning – Indicates that the voltage is close to the limit of the allowable range.
• Error – Indicates that the voltage is outside the allowable range.
Lower Limit Warnings Number of low limit voltage warnings.
Upper Limit Warnings Number of high limit voltage warnings.
Lower Limit Errors Number of low limit voltage errors.
Upper Limit Errors Number of high limit voltage errors.