GERAFT LOGIQ P9/P7
D
IRECTION 5604324, REVISION 11 DRAFT (JANUARY 24, 2019) SERVICE MANUAL
5-70 Section 5-11 - Common Service Desktop (R3 only)
This table shows all the elements available on DICOM Service Status with descriptions.
For more information, see:
DICOM on page 5-69
DICOM Spooler on page 5-71
Table 5-28 DICOM Service Status
Element DESCRIPTION
DICOM Details
AE Title of Console Application entity title of the LOGIQ P9/P7.
Port No of Console Port number of the LOGIQ P9/P7.
DICOM Connectivity Status
Service Name Name of the DICOM services configured for the LOGIQ P9/P7.
Status
Status of the DICOM service. Valid values are:
• OK
• Not Ok
• Not Ping
• Ping in progress
• Queued
Error Details Details for any errors associated DICOM services.
Type Type of error.
Device Name Name of the device.
AE Title Application entity title for the device.
IP Address IP address for the device.
Port Port for the device.
Ping Device Select to confirm that a device is connected or not.