GEDRAFT LOGIQ P9/P7
D
IRECTION 5604324, REVISION 11 DRAFT (JANUARY 24, 2019) SERVICE MANUAL
Chapter 5 - Components and Functions (Theory) 5-77
Disable Remote Connectivity
This utility will uninstall remote connectivity items like Agent Configuration, SSH and Virtual Console
Observation (VCO) from Service Desktop. If remote connectivity is disabled, then the Submit Changes
button on Agent Configuration and the Start/Stop buttons for SSH and VCO will all be disabled. In
addition, some scripts will execute to make the LOGIQ E10 more secure.
NOTE: If you have disabled remote connectivity and then performed a software reload, you will have
to disable remote connectivity again.
To access Disable Remote Connectivity, navigate to Utility (second page) > Service > Utilities, and then
select Disable Remote Connectivity.
Figure 5-52 Disable Remote Connectivity
To disable remote connectivity:
1.) Navigate to Utility (second page) > Service > Utilities, and then select Disable Remote Connectivity.
2.) Select Disable Remote Connectivity.
3.) To disable connectivity to the LOGIQ E10 through Agent Configuration, SSH, and Virtual Console
Observation, click OK.
4.) Once connectivity has been disabled, select OK again.

Table of Contents

Other manuals for GE LOGIQ P7

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ P7 and is the answer not in the manual?

GE LOGIQ P7 Specifications

General IconGeneral
BrandGE
ModelLOGIQ P7
CategoryMedical Equipment
LanguageEnglish

Related product manuals